Describe
TLC9803 dynamic ECG systems adapts 3-lead system defined by our company, can continuously record ECG waves for 24h and analyze waves with PC software. It is applicable in hospital and community medical institution.
Recorder Features
With advanced design and technology, the recorder can work stably, reliably and durably.
Be provided with the ability of anti-jamming and aseismatic.
Small volume with OLED screen. The resolution is 160*128.
Be provided with functions of waveform preview, record review and event marking.
Accurately record the start time of sampled data due to the real-time clock of itself.
Avoid damage caused by repeated insertion and extraction by using the built-in TF card to store the data.
The maximum capacity can be 2gb, so that the original data can be perfectly saved by no compression.
The average review time of individual cases can be less than 40 seconds by using the interface of USB 2.0.
Record waveform details inextenso by the design of high precision and sampling frequency.
Record the state of pacemaker accurately by higher sampling frequency.
The features of software
Fast and accurate analysis system.
The 3-lead synchro analysis to find accurate QR can be accurate and no distortion.
There are more than 10 templates as atrial early beat module, ventricular early beat module, long interval module, atrial flutter module, atrial fibrillation module etc. and multiplied user-defined templates could identify almost any kind of pathological waveforms.
Flexible analysis channel selection function, can choose any channel as the main analysis of the channel.
Flexible atrial fibrillation analysis, so that doctors can use full, segmented automatic, manual analysis of atrial fibrillation, make the analysis faster and more accurate.
Powerful pacemaker analyze, on the aai, vvi, ddd etc. all pacemaker analyze.
Many pacing modules have been added.
Fast review analysis function, each period of review can be single lead or all leads electrocardiogram.
Analysis of hearing rate variability with short range 5 minutes and long range 1 hour and 24 hours analysis of heart rate variability.
One-stop printing operation, report printing process is convenient and fast.
Perfect function of case management.
Advanced ST segment analysis and myocardial ischemia total strain, can for the ST abnormal analysis according to the whole and event, make the doctor can comprehensive judgment myocardial ischemia.
Unique "sleep breath pause syndrome analysis" analysis can predict sleep breath pause risk.
The "hrt" analysis can predict the death risk in patients with myocardial infarction.
The "t wave alternation" analysis is an important index to predict malignant arrhythmia and sudden cardiac death.
Specification
Number of channels: 3 channels
Sampling accuracy: 12bit
Record time: 24h
Power: 2 AA size batteries
Interface: USB 2.0
Scale voltage: 1mV ± 5%
Standard sensitivity: 10mm/mV ± 5%
Noise level: ≤ 30μV
CMR: ≥ 60dB
Low frequency characteristics: Time constant ≥ 3.2s
Scan speed: 25mm/s ± 5%
Permanent polarization voltage: ± 300mV DC polarization voltage, sensitivity shift ≤ ± 10%
Minimum measurement signal: 50 µS v p-p
Product safety type: Type B (internally powered)
Packing list
Monitoring tester
3-lead ECG cable (1 set)
ECG electrode 1 bag (20 per bag)
USB data cable (1)
Disk (1)
Case (1)