Introduction
SP950 is a syringe pump for medical use and adopts micro computer system and micro electronics
technology, features in high intelligence and high safety. LCD can display every parameter in real time,
With multifunction salarm and alarm with visual, auditory and character prompt. It takes the advantages
of easy operation, steady work, high accuracy and checking syringe specification automatically.
And now it has been widely applied in every vein injection in hospitals for clinical nursing, such as ICU.
CCU, Neonatology, Obstetrics and Gynecology, Pediatrics, Internal Medicine, Surgery, Operating Room,
Emergency Room.
Functions
1) The infusion rate and total volume can be set by user.
2) Alarms for infusion tube occlusion, syringe fall-off, communication abnormal, low battery.
3) Bright, alarm volume and range of pressure alarm are adjustable.
4) Fast feed function, convenient for delivering medicine quickly and immediately.
5) Enter KVO mode after infusion is completed.
6) The system will give an immediate signal if no operation in power-on state.
7) Built-in rechargeable battery provides such functions as auto-switch power supply and auto-charge.
8) Display parameters with large screen and give quick information with visual, auditory and character
signal, which is convenient for medical persons to operate.
9) The device can be equipped with disposable syringe (CNS), and accuracy can be ensured by setting
parameters.
Performance
1) Specification requirement for syringe: 10ml/20ml/30ml/50ml (cns)
2) Infusion rate:
10ml 0.1ml/h-100ml/h
20ml 0.1ml/h-200ml/h
30ml 0.1ml/h-300ml/h
50ml 0.1ml/h-500ml/h
3) Rate of rapid feeding:
10ml 100ml/h
20ml 200ml/h
30ml 300ml/h
50ml 500ml/h
4) Accumulated volume display: 0ml-999.9ml
5) Accumulation: ±3%
6) Pressure range for occlusion salarm:
Three grades are adjustable:
Low: 40kpa ±16kpa
Middle: 70kPa ± 30kPa
High: 100kPa ± 35kPa
7) KVO rate: 0.1ml/h ~ 5ml/h, adjustable
8) Safety classification: Class, equipment, internally driven type BF
9) Alarm function: 1 in non-operating state 2 near empty 3 empty, enter KVO mode 4 syringe fall off 5
occlusion 6 pressure failure 7 motor failure 8 low battery 9 communication abnormal
10) Maximum power consumption: 20VA ~ 30VA, the device can work continuously for at least two hours
With the injection rate of 5ml/h or 500ml/h after fully charged.
11) Power supply: AC100~240V 50/60Hz
12) Battery: Rechargeable lithium battery, 7.4V, 3500mAh
13) Working environment: Temperature: +5~40℃; Atmospheric pressure: 860hpa~1060h; Relative humidity
: 20%~90%; no strong vibration and no corrosive gas.
14) Storage environment: Temperature: -20℃~+55℃; Relative humidity: ≤90%, no condensation; no corrosive gas and well ventilated
Accessories
One bracket
One power cord
One user manual
Two fuses
Physical property
Dimension: 310mm * 140mm * 135mm (l×w×h)
Weight: 1.7kg